Almost all STMs use piezo translators to scan the tip, or seldom, to scan the sample. Even the first STM[84,85] and some of the predecessor instruments[86,87] used piezo translators for scanning. Microscopes using magnetostrictive materials[88] or electromagnetic drives[89] have been proposed. We will concentrate on piezo electric materials.
An electric field applied across a piezo electric material causes a change in
the crystal structure, with expansion in some directions and contraction in
others. Also, a net change in volume occurs. Detailed descriptions of the piezo
electric effect can be found in solid state physics
textbooks[90]. The transverse piezo electric effect is by
far the most important for scanning probe microscopes. The expansion
perpendicular to the applied electric field
for a long slab of
material with the field applied across the small sides is
![]() |
(931) |
where
is the piezoelectric constant,
the applied voltage and
the thickness of the piezo slab or the
distance between the electrodes where the voltage is applied.
This allows to choose the sensitivity of a piezo actuator within the limits
of its mechanical stability.
![]() |
The first STMs all used piezo tripods for scanning (see for instance Binnig and
Rohrer[45]). The piezo tripod (figure
4.273a)) is an intuitive way to generate the three dimensional
movement of a tip attached to its center. However, to get a suitable stability
and scanning range, the tripod needs to be fairly large (about 5 cm). Its size
and its asymmetric shape make it very susceptible to thermal drift. The design
of van Kempen and van de Walle[82] (figure
4.273b)) tries to circumvent this problem by using a
symmetrical design. Its thermal drift performance is much better than the
simple tripod. However a complicated assembly of many piezo pieces is required.
The tube scanner (figure 4.273c)) is now widely used in
scanning tunneling and scanning probe microscopy for its simplicity and its
small size[91]. The outer electrode is segmented in four
equal sectors of 90 degrees. Opposite sectors are driven by signals of the same
magnitude, but opposite sign. This gives, through bending, a two dimensional
movement on, approximately, a sphere. The inner electrode is normally driven by
the
signal. It is possible, however, to use only the outer
electrodes for scanning and for the
-movement. The main drawback of applying
the
-signal to the outer electrodes is, that the applied voltage is
the sum of both the
- or
-movement and the z-movement. Hence a larger
scan size effectively reduces the available range for the
-control.
Piezo scanners, tubes and tripods, are made of piezo ceramic material.
Piezo materials with a high conversion ratio, i.e., a large
or small
distances between the electrodes,
allowing large scan ranges with low driving voltages, do have substantial
hysteresis resulting in a deviation from linearity by more
than 10 %. The sensitivity of the piezo ceramic material (mechanical
displacement divided by driving voltage) increases with reduced scanning
range, whereas
the hysteresis is reduced. A careful selection of the material for the piezo scanners,
the design of the scanners, and of the operating conditions is necessary to
get optimum performance.
Copyright by Othmar Marti and Alfred Plettl, 2007-08-14